Evaluación del uso de trampa de evaporación en la evaporación en la calibración de pipetas de pistón en volúmenes menores o iguales a 100 µL
This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation....
Guardat en:
| Autor principal: | |
|---|---|
| Format: | Online |
| Idioma: | spa |
| Publicat: |
Universidad de Costa Rica
2017
|
| Matèries: | |
| Accés en línia: | https://revistas.ucr.ac.cr/index.php/ingenieria/article/view/28200 |
| Etiquetes: |
Afegir etiqueta
Sense etiquetes, Sigues el primer a etiquetar aquest registre!
|
| Sumari: | This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation. 10 repetitions were performed at three different points and the values provided by the calibration certificate of the analyzed pipette were compared to a reference value determined by the National Metrology Laboratory (LACOMET). |
|---|