Evaluación del uso de trampa de evaporación en la evaporación en la calibración de pipetas de pistón en volúmenes menores o iguales a 100 µL

This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation....

Descripció completa

Guardat en:
Dades bibliogràfiques
Autor principal: Méndez Arias, Johanna
Format: Online
Idioma:spa
Publicat: Universidad de Costa Rica 2017
Matèries:
Accés en línia:https://revistas.ucr.ac.cr/index.php/ingenieria/article/view/28200
Etiquetes: Afegir etiqueta
Sense etiquetes, Sigues el primer a etiquetar aquest registre!
Descripció
Sumari:This paper aims to demonstrate the importance of using evaporation trap during piston pipette calibration, specifically at microliter levels (μL). Calibrations were performed in a laboratory pipette using the gravimeter method with evaporation trap, and without trap or any correction by evaporation. 10 repetitions were performed at three different points and the values provided by the calibration certificate of the analyzed pipette were compared to a reference value determined by the National Metrology Laboratory (LACOMET).